Principle of xray collimator simulating illumination field

There is an analog light source in the xray collimator to simulate the X-ray field.

At present, there are two kinds of analog light sources in the market, one is LED lamp, the other is halogen lamp.

Principle of simulated illumination field: If simulated illumination field can correctly reflect X-ray, the position of X-ray focus, point source and reflector is very important.

The position relations of X-ray focus, point source and reflector are as follows:

The position of the point source and the X-ray focus must be symmetrical with the mirror as the symmetrical central axis.

Reflector is a very important part in xray collimator, because its position will directly affect whether the simulated illumination field is correct.

collimated beam